THOR PWG-33 Pulmonary Waveform Generator

, ,

General description

THOR PWG-33 provides waveforms with pre-set volume and flow values for calibrating respiratory monitoring apparatus and other equipment measuring flow and volume. Waveforms generated by PWG-33 and the evaluations of test results are fully compatible with the American Thoracic Society specifications released on 11 November 1994 and the EN 13826:2003 European Standard.
ISO 9001 and ISO 13485 Quality Management System guarantees the reliability of development and manufacturing. All the parameters of generated waveforms are traceable to National Standards.

Generated wavefront

  • 24 ATS Standard Volume(time) waveforms
  • 26 ATS Standard Flow(time) wavefront
  • 13 Standard waveforms according to EN ISO 26782:2009 Annex C
  • 10 Standard waveforms according to EN ISO 23747:2009 Annex C, profile A
  • User configured waveforms according to EN ISO 23747:2009 Annex C, profile B

Main features

  • BTPS simulation with heated and humidified air for all ATS standard waveforms
  • Full automatic measurement of flow meter’s resistance
  • Full support of human differential tests, according to ATS
  • Evaluation of test results and error analysis, according to ATS
  • Full automatic test user definable sequence
  • Printed test report
DSCN0005.JPG

Block diagram

Block_diagram.jpg

Item’s description

Item Description
1 Calibration ports
2 Control button
3 Remote control port
4 USB plug
5 Pneumatic outlet
6 Humidifier connected to the BTPS inlet
7 Two way (BTPS) valve
8 Heated syringe
9 Circuit board
9a Main circuit with the internal microcontroller
9b Separate calibration circuit
10 Scheider Electric SD326 stepper motor controller
11 Berger-Lahr stepper motor
12 Photo interrupters: end-position sensors
13 Photo interrupters for kinematic test
14 Timing belt driving the piston

Specification

PNEUMATIC INFORMATION
Pnaumatic outlet ISO-30 Medical taper Male OD=30,2 mm Basic taper 1:20
Maximal back-pressure of the object being tested 4.0 kPa at 12 l/s
Volume resolution 0.345 ml
Maximal volume 10.00 l
Maximal flow rate 20.0 l/s
Minimal rise time to flow rate 16 l/s at black-pressure 3 kPa 4 ms
Maximal slew rate to flow rate 16 l/s at back-pressure 3 kPa 4000 l/s2
ACCURACY
Volume 0.2 % or 10 ml which ever is larger
Flow 0.3 % or 15 ml/s which ever is larger
Time 0.02 % or 1 ms which ever is larger
Temperature of air with BTPS simulation 37 °C ± 1 °C
Relative humidity of air with BTPS simulation 92 % ± 5 %
Distance between photo interrupters for verification (nominal value) 127 mm
TRANSPORTING CONDITIONS
Temperature -30 °C ÷ +50 °C
Relative humidity 10 % ÷ 100 % (non condensing)
Ambient pressure 500 ÷ 1060 mbar
STORAGE CONDITIONS
Temperature 0 °C ÷ +50 °C
Relative humidity 10 % ÷ 85 %
Ambient pressure 500 ÷ 1060 mbar
OPERATING CONDITIONS
Temperature +10 °C ÷ +35 °C
Relative humidity 30 % ÷ 75 %
Ambient pressure 700 ÷ 1060 mbar
ELECTRICAL DATA
Nominal voltage 115 or 230 VAC
Mains voltage tolerance at 115 VAC 100 ÷ 130 VAC
Mains voltage tolerance at 230 VAC 170 ÷ 240 VAC
Frequency 50 / 60 Hz
Power consumption; PWG-33 max. 25 VA
Power consumption; PWG-33/BT max. 770 VA
MECHANICAL PROPERTIES
Size without humidifier 1100 mm x 370 mm x 265 mm
Weight 38 kg
MINIMAL PC CONFIGURATION
Operating system Windows XP, Vista 7
Computing capacity See operating system recommendations
Platform booth 32 and 64 bit supported
Free USB 1.1 or 2.0 compatible port 1 port