OxiThor – USB Oximeter
General Description
With USB interface, the product can connect with PC. The PC software is compatible with other oximeter product of CMS system. With high precision measuring circuit,the measuring result is exact and credible. Light weight, small size, easy to carry.
Specification
The Pulse Oxygen Saturation (SpO2) Measuring
- Measuring range:0% – 100%
- Accuracy:70%-99%:+/-2% ;Below 70% unspecified.
Pulse Rate Measuring
- Measuring range:25bpm – 250bpm
- Accuracy: +/-2bpm or +/-2% (select larger)
- SpO2 is 1%; PR is 1bpm.
Measurement Performance in Weak Filling Condition
- SpO2 and pulse rate can be shown correctly when pulse-filling ratio is 0.4%. SpO2 error is +/-4%;pulse rate error is +/-2 bpm or +/-2% (select larger).
Resistance to surrounding light
- The deviation between the value measured in the condition of man-made light or indoor natural light and that of darkroom is less than +/-1%.
Work Environment Requirement
- Environment temperature: 0°C – 50°C
- Relative humidity: 15% – 95%
- Atmospheric pressure: 700hPa – 1060hPa